Custom CT protocol exposes Cedars-Sinai patients to excessive dose

By James Brice | October 15, 2009

Cedars-Sinai Medical Center, the prestigious Los Angeles healthcare institution known as the hospital to the Hollywood stars, has been jolted by an FDA alert indicating that perfusion CT performed during an 18-month period exposed more than 200 stroke patients to eight times the normal dose of ionizing radiation for the procedures.

A custom perfusion CT protocol developed at Cedars-Sinai is considered the prime suspect for elevated doses discovered in August when a patient complained of patchy hair loss -- a cardinal symptom of radiation poisoning -- after receiving a CT procedure.

For more on this topic:
CT angiography data offer new approach to perfusion stroke imaging

CT perfusion expands stroke care options

Perfusion adds clinical value to CT evaluation of stroke

According to an Oct. 8 FDA alert, 206 patients were exposed to elevated doses for perfusion CT ordered to characterize ischemic stroke. Patients received doses of 3 Gy to 4 Gy instead of the standard 0.5 Gy for the procedure. When notified, eight additional patients reported hair loss and skin reddening. The symptoms appeared within six weeks of scanning, according to the hospital. Cedars-Sinai reported the incident to the California Department of Public Health.

Emergent perfusion CT is typically performed on stroke patients to select candidates for endovascular or intra-arterial clot removal therapy during the hours after symptom onset, according to Dr. Michael Lev, director of emergency neuroradiology at Massachusetts General Hospital.

"It identifies anatomic regions at risk and how complete the infarction is at the time of admission," he said in an interview.

Standard perfusion CT for stroke involves 80 kV and 200 mAs scans, according to Lev, a well-known developer of perfusion protocols. The typical 45 to 60-sec acquisition exposes the patient 2 to 3 mSv, about the same amount of radiation as a routine CT head scan, though concentrated in a smaller area, he said.

E-mail traffic among radiologists had them speculating that the Cedars-Sinai group was using 120 kV or 140 kV settings. A lower kV setting would have increased the conspicuity of the iodine(Drug information on iodine) in the contrast media but would have otherwise produced noisier images, Lev said. The mA setting was also apparently modified.

"They may have tried to use a higher mA setting to deliver more photons for less noisy images," he said, "But with CT perfusion what you really want to optimize is the iodine conspicuity."

The procedures were performed on a GE multislice scanner. The specific model number was not disclosed. In a written statement, GE Healthcare said it is cooperating with the FDA to investigate the incidents. It said no malfunctions or defects in any of the GE scanners used for perfusion CT at Cedars-Sinai were found.

Though the specific incident may be unique to Cedars-Sinai, the FDA characterized the magnitude of the overdoses and their impact on affected patients as "significant." During its initial assessment, the FDA identified shortcomings with the CT quality assurance program at the Los Angeles hospital that could arise at other hospitals and diagnostic imaging procedures. Such incidents could go undetected for the procedures, exposing patients to a level of radiation high enough to elevate their future health risks while not high enough to induce immediate symptoms, the FDA said.

In a written statement, Cedars-Sinai acknowledged that its CT perfusion protocol was modified in February 2008 to improve the quality of data for stroke diagnosis. Patients were exposed to elevated radiation doses because of "a misunderstanding about the embedded default setting applied by the machine in this brain perfusion scan, which is different from a regular CT brain scan."

Cedars-Sinai instituted additional "double-checks" in its operation of the CT scanner and additional equipment protocols to ensure that the problem is not repeated, according to the statement.

The FDA encouraged every facility performing CT imaging to review its protocols and to be aware of the dose indices normally displayed on the control panel before scanning. Key indices include the volume CT dose index (CTDIvol) and the dose-length product (DLP).

The agency urged users to carefully monitor the dose indices displayed on the control panel for each protocol selected before patient scanning. To prevent accidental overexposure, they were asked to make sure that the values displayed reasonably correspond to the doses normally associated with the protocol. The displayed values should be confirmed again after the patient has been scanned.

"The take home message is that if you do perfusion CT for stroke using the standard accepted parameters for mA and kV and deciding against doing serial scans, especially for patients who are receiving other CT and CTA studies, you should be safe relative to the risk of mismanaging these acute stroke patients," Lev said.

 

Join the Conversation

Want to join the conversation? Just sign in or register today to become part of our growing, online community.

Computer-aided detection software backs up coronary CTA interpretation

Custom CT protocol exposes Cedars-Sinai patients to excessive dose

CTA beats calcium score in predicting near-term coronary events

Cardiac CT exhibits both clinical and cost benefits

Informatics investments prove vital in competitive world

Computer-aided detection software backs up coronary CTA interpretation






TopicIndex

 

ACOs
Cardiac
Case Studies
Colonography
CT
Digital X-ray
Direct Radiography
Elastography
Low-Dose Modalities
Meaningful Use
Molecular Imaging
MRI
 

 

Nuclear
PACS
PET/CT
PET/MR
Practice Management
RIS
Teleradiology
Ultrasound Imaging
Vendors
Voice Recognition
Women's Imaging
All Topics
 


SponsoredResources


OptumInsight
Acadiana Computer Systems, Inc. gains a 100% ROI on their radiology billing


Key Equipment Finance
Michiana Hematology Oncology Success Story


Barco
Multi-modality breast imaging using RapidFrame™ technology


Siemens
3D Ultrasound of the Breast


Ziosoft, Inc.
PhyZiodynamic Solutions: Applying Supercomputing to Patient Care


Siemens
Easy Guide to Low Dose


Medrad
Improving Clinical Outcomes and Workflow
Toshiba America Medical Systems
Minimizing dose, sedation in pediatric CT

 

View All

 


FromPhysiciansPractice

'What They Should Really Teach in Medical School'
Julie Schopps, MD , February 6, 2012
The North Carolina-based pediatrician weighs in on why she thinks the real learning doesn't take place until students are out of the classroom.
Improve EHR Systems by Rethinking Medical Billing
Daniel Essin, MA, MD, February 6, 2012
Separating billing-related data from other clinical documentation and transmitting it to a billing system is not difficult …no matter how the charting is done.
Keeping Your Medical Practice’s Accounts Receivable on Track
P.J. Cloud-Moulds, February 4, 2012
Here are the minimum reports you should be running to keep an eye on your practices A/R.
Healthcare Providers Play Crucial Role in Helping Victims of Abuse
Stephen Hanson, PA-C , February 3, 2012
I would urge each and every one of you to be familiar with the warning signs of abuse, and the resources available to you all as healthcare providers.
Protecting Your Medical Practice's Data
Marisa Torrieri, February 3, 2012
Here's the scoop on how to implement a good data-backup plan at your office.
  • On This Site
  • Most Emailed
  • On This Topic

MostPopular

  • Whole-breast ultrasound brings significant screening benefits

    JAN 15 2010 DIAGNOSTIC IMAGING ASIA PACIFIC READ >>

  • CNN Investigation Targets Radiology Board Exam Cheating

    JAN 13 2012 READ >>

  • As teleradiology evolves, it changes dramatically, plays growing role in practice

    DEC 15 2010 DIAGNOSTIC IMAGING READ >>

  • Delayed side effects persist in IV iodinated contrast media

    MAY 28 2009 DIAGNOSTIC IMAGING EUROPE READ >>

  • Mucinous Adenocarcinoma of Stomach

    JAN 9 2012 READ >>

MostPopular

  • CNN Investigation Targets Radiology Board Exam Cheating

    JAN 13 2012 READ >>

  • Telemammography Taking Hold

    JAN 24 2012 READ >>

  • Riverain’s Chest X-Ray Comparison Tool Gets FDA Nod

    JAN 11 2012 READ >>

  • Podcast: Implementing a Hybrid PET/MR System

    JAN 30 2012 READ >>

  • Taking Medical Image Sharing to the Cloud

    JAN 19 2012 READ >>

MostPopular

  • Whole-breast ultrasound brings significant screening benefits

    JAN 15 2010 DIAGNOSTIC IMAGING ASIA PACIFIC READ >>

  • CNN Investigation Targets Radiology Board Exam Cheating

    JAN 13 2012 READ >>

  • As teleradiology evolves, it changes dramatically, plays growing role in practice

    DEC 15 2010 DIAGNOSTIC IMAGING READ >>

  • Delayed side effects persist in IV iodinated contrast media

    MAY 28 2009 DIAGNOSTIC IMAGING EUROPE READ >>

  • Mucinous Adenocarcinoma of Stomach

    JAN 9 2012 READ >>

  • Popular
  • Recent

Comments

  • Poll of the Week: Is the Use of Recalls Cheating?

    JAN 26 2012 READ >>

  • CNN Investigation Targets Radiology Board Exam Cheating

    JAN 13 2012 READ >>

  • Columbus Radiology Launches Imaging Ordering App

    JAN 19 2012 READ >>

  • Radiology Comic: Doctors Cheating

    JAN 31 2012 READ >>

  • New MRI Algorithm Cuts Scan Time by Two-Thirds

    DEC 12 2011 READ >>

Comments

  • Plasma Cell Myeloma

    FEB 1 2012 READ >>

  • PET/MR Machines – A Comparison

    JUN 13 2011 DIAGNOSTICIMAGING.COM READ >>

  • Reporting Law Prompting Greater Attention to Radiation Dose

    FEB 6 2012 READ >>

  • Poll of the Week: Do You Deliver Imaging Results Directly to Patients?

    FEB 2 2012 READ >>

  • Delayed side effects persist in IV iodinated contrast media

    MAY 28 2009 DIAGNOSTIC IMAGING EUROPE READ >>

JobListings

Post a job

Powered by SearchMedica Jobs


SearchMedicaSearchResult

Find peer-reviewed literature and websites for practicing medical professionals

CME on Informatics Pacs
Evidence on Informatics Pacs
Guidelines on Informatics Pacs
Patient Education on Informatics Pacs
Clinical Trials on Informatics Pacs
Practical Articles on Informatics Pacs
Research and Reviews on Informatics Pacs
All "Informatics Pacs" results

FeaturedWhitePaper


More white papers >>


CancerNetwork | CME LLC | ConsultantLive | Diagnostic Imaging | Musculoskeletal Network | OBGYN.net | PediatricsConsultantLive |
Physicians Practice | Psychiatric Times | SearchMedica | Medical Resources

© 1996 - 2012 UBM Medica LLC, a UBM company
Privacy Statement - Terms of Service - Advertising Information - Editorial Policy Statement - UBM Medica Network Privacy Policy