External fixation devices and MRI safety

January 17, 2006

Frank Shellock, Ph.D.

Adjunct clinical professor of radiology and medicine, University of Southern California and Institute for Magnetic Resonance Safety, Education, and Research

www.MRIsafety.com

www.IMRSER.org

Most orthopedic implants and materials do not pose problems for patients undergoing MRI procedures. MRI may be hazardous for external fixation systems, however, because of the length of the implant or the formation of a conductive loop.

External fixation systems are specially designed frames, clamps, rods, rod-to-rod couplings, pins, posts, fasteners, wire fixations, fixation bolts, washers, nuts, hinges, sockets, connecting bars, screws, and other components used in orthopedic and reconstructive surgery.

Indications for external fixation systems are varied, including the following treatment applications:

  • open and closed fracture fixation

  • pseudoarthroses of long bones (congenital and acquired)

  • limb lengthening by metaphyseal or epiphyseal distraction

  • correction of bony or soft-tissue defects

  • correction of bony or soft-tissue deformities

The assessment of MRI safety issues for external fixation systems is especially challenging because of the myriad possible components, many of which are made from conductive materials, and the many configurations used for these devices.

The primary concern is MRI-related heating, which is dependent on the particular aspects of the external fixation system. The specific MRI conditions — strength of the static magnetic field, radiofrequency, type of RF transmit coil, pulse sequence, body part imaged — directly affect the safety aspects of scanning patients with external fixation systems.

For each external fixation device that has undergone MRI safety testing and applied for approval of labeling from the FDA, highly specific guidelines need to be followed relative to a given device configuration and the MRI conditions used for the imaging procedure. This information will be contained in the Instructions for Use for a given external fixation device. Imaging protocols will vary accordingly.

REFERENCE

Shellock FG. Reference manual for magnetic resonance safety, implants, and devices. 2006 ed.

 

Join the Conversation

Want to join the conversation? Just sign in or register today to become part of our growing, online community.





Article Tools
Sponsored White Papers


SponsoredResources


OptumInsight
Acadiana Computer Systems, Inc. gains a 100% ROI on their radiology billing


Key Equipment Finance
Michiana Hematology Oncology Success Story


Barco
Multi-modality breast imaging using RapidFrame™ technology


Siemens
3D Ultrasound of the Breast


Ziosoft, Inc.
PhyZiodynamic Solutions: Applying Supercomputing to Patient Care


Siemens
Easy Guide to Low Dose


Medrad
Improving Clinical Outcomes and Workflow
Toshiba America Medical Systems
Minimizing dose, sedation in pediatric CT

 

View All

 
What's New on DiagnosticImaging.com

So, Imaging Technologist, You Want to Be Considered a Medical Professional
February 10, 2012
MRI Technique Shows Success of MS Drug Copaxone
February 9, 2012
Early CT Angiography Identifies Recurrent Stroke Risk
February 9, 2012
Podcast: Using MRI in the Operating Room
February 8, 2012
PET with FDG May Predict Outcomes in Triple-Negative Breast Cancer
February 8, 2012
Featured White Paper


More white papers >>



CancerNetwork | CME LLC | ConsultantLive | Diagnostic Imaging | Musculoskeletal Network | OBGYN.net | PediatricsConsultantLive |
Physicians Practice | Psychiatric Times | SearchMedica | Medical Resources

© 1996 - 2012 UBM Medica LLC, a UBM company
Privacy Statement - Terms of Service - Advertising Information - Editorial Policy Statement - UBM Medica Network Privacy Policy