Semiautomated vessel analysis software offered by five vendors tends to underestimate the degree of carotid stenosis, particularly at the higher levels that indicate a need for corrective action, according to a study presented at the RSNA meeting.
Semiautomated vessel analysis software offered by five vendors tends to underestimate the degree of carotid stenosis, particularly at the higher levels that indicate a need for corrective action, according to a study presented at the RSNA meeting.
Among the findings of a research team led by Dr. William Boonn: Underestimation of stenosis greater than 70% ranged from 12% to 50%. Boonn, an imaging informatics fellow at VA Maryland Health Care System when the study was conducted, is now at the University of Pennsylvania.
Using semiautomated vessel analysis software led to significant variability and underestimation of carotid stenosis, the study concluded. These results, along with analysis failures, worsen as the percentage of stenosis increases.
Precise data for individual workstations and software packages were not disclosed during the presentation.
The study analyzed data from 23 carotid arteries from 12 patients undergoing carotid CT angiography using multislice CT. One artery was excluded as a result of postsurgical changes. The software measurements were compared with manual measurement of all 23 carotid arteries performed in consensus with a standard procedure described in recent literature and based on criteria in the North American Symptomatic Carotid Endarterectomy Trial (NASCET).
Users of semiautomated vessel analysis systems must select the vessel to be segmented and the segment to be analyzed, Boonn said. Automated results from six commercially available systems and packages from five vendors were compared with the manual measurements.
No variation in the manual and automated measurements appeared among the eight normals in the study group, Boonn said. But as the degree of stenosis increased, the variability increased as well, mostly in the form of underestimation.
This finding is critical, because the NASCET has shown that patients with high-grade stenosis can benefit from carotid endarterectomy.
Even with the underestimation of stenosis, vessel analysis systems offer some benefits, Boonn said. Segmentation can improve visualization, and curved multiplanar reformatting can help visualize tortuous vessels.
But the study's results suggest that vendors must improve the consistency and accuracy of automated stenosis measurements, according to Boonn.
"We advocate the use of publicly available standardized data sets that can be used by vendors to validate their measurements," he said.
For more online information, visit Diagnostic Imaging's RSNA Webcast.
The Reading Room Podcast: Current Perspectives on the Updated Appropriate Use Criteria for Brain PET
March 18th 2025In a new podcast, Satoshi Minoshima, M.D., Ph.D., and James Williams, Ph.D., share their insights on the recently updated appropriate use criteria for amyloid PET and tau PET in patients with mild cognitive impairment.
Study with CT Data Suggests Women with PE Have More Than Triple the One-Year Mortality Rate than Men
April 3rd 2025After a multivariable assessment including age and comorbidities, women with pulmonary embolism (PE) had a 48 percent higher risk of one-year mortality than men with PE, according to a new study involving over 33,000 patients.